Ozone Monitor Models 106-L, 106-M, 106-MH and 106-H
The Model 106 series was designed as an "ozone monitor on a board" in which nearly all of the components are mounted directly to the printed circuit board. Thus, the Model 106 may be purchased without the enclosure for OEM applications. See: Model OEM-106.
The Model 106-L is designed for ambient low measurements down to very low ppb levels and has been designated by the EPA as a Federal Equivalent Method (FEM) for monitoring for compliance with the US Clean Air Act: EQOA–0914–218.
Specifications for Different Optical Benches
106-L | 0-100 ppm | 0.0001 ppm (0.1 ppb) | Higher of 0.0015 ppm (1.5 ppb) or 2% of reading |
106-M | 0-1,000 ppm | 0.01 ppm | Higher of 0.01 ppm or 2% of reading |
106-MH | 0-10,000 ppm (0-1 vol%) | 0.1 ppm | Higher of 0.05 ppm or 2% of reading |
106-H | 0-20 wt% O2 or Air 0-14 vol% | 0.01 wt% 0.01 vol% | Higher of 0.01 wt% or 2% of reading |
General Specifications
Measurement Principle | UV Absorption at 254 nm |
Measurement Interval | 2 s |
Data Averaging Options | 10 s, 1 min, 5 min, 1 hr |
Flow Rate | ~1 Liter/min; Up to 20 L/min for Model 106-H |
Data Storage | 32,736 lines (10 s avg. = 3.7 days; 5 min avg = 113 days) |
Choice of Units | ppb, pphm, ppm, μg m-3, mg m-3 (for Models 106-L and 106-M) ppm, mg m-3 (for Model 106-MH) vol%, wt% air, wt% O2, g/m3, g/Nm3 (for Model 106-H) |
Data Outputs | USB, RS232, 0-2.5 V Analog, 4-20 mA, LCD Display |
Power Requirements | 11-28 V DC, nominally 500 mA at 12 V DC, 6 watt |
Size | 3.6 x 7.9 x 9.4 inches (9 x 20 x 24 cm) |
Weight | 4.2 lb (1.9 kg) |
Features
- Measurement based on UV absorption
- Low power consumption; can be battery operated (optional lithium-ion battery)
- Internal data logger with real-time clock
- On-board microprocessor with interactive menus includes data averaging options of 10 s, 1 m, 5 m and 1 hr
- USB and RS-232 output of time/date, O3 concentration, internal temperature and pressure
- Analog output (0-2.5 V and 4-20 mA) of ozone concentration in user selected units
- Two 2-level relays for control purposes (e.g., control of ozone source or turn warning light on and off)
- Long-life pump (15,000 hr) for Models 106-L, 106-M and 106-MH
- Bluetooth option for wireless data transmission